The table contains all information about this Round Robin BOX. Please check the table frequently. For each intercomparison fill out the form bellow.
I will check and place all on the table. After analysing the data you will find the results under "Results" in the table for each box and the whole report on the ECAC website.
Start intercomparison | Station | End intercomparison | Information |
Friday March 23 2018 | Granada (E) | Wednesday April 11 2018 | march 15, 2018; CPC arrived on the station inter-comparison finished Plots are on the website and Report is done |
Tuesday April 17 2018 | Arenosillo (E) | Wednesday May 16 2018 | inter-comparison finished Plots are on the website and Report is done |
Tuesday October 04 2018 | Montseny (E) | Wednesday October 10 2018 | May: The MPSS is for maintenance in TROPOS -> intercomparison after Izana September 27: CPC is traveling from Izana to Montseny Oct 05 -> Box arrived in Montseny but it was damaged. E-mail from 05.10.2018 Natalia to Noemi Oct 05 -> Noemí will setup the TCPC on station from Oct 24 -> inter-comparison on station inter-comparison finished |
Madrid (E) | Arenosillo will send the TCPC to Madrid May 22 -> TCPC arrived in Madrid - inter-comparison from June 01 -waiting for data and CPC infomration -> July 09 | ||
Friday August 24 2018 | Izana (E)* | Wednesday September 12 2018 | Madrid will send the TCPC to Izana -> July 09 Izana send all the data September 17 -> TROPOS will check the all |
Station | Address | Contact Person |
Granada (E) | Instituto Interuniversitario de Investigación del Sistema Tierra en Andalucía(IISTA-CEAMA), Av. del Mediterraneo SN | Juan Andrés Casquero e-mail: casquero@ugr.es Phone: |
Arenosillo (E) | National Institute for Aerospace Technology (INTA) El Arenosillo - Atmospheric Sounding Station Crta. San Juan del Puerto-Matalascañas, km.34 21130, Mazagón - Huelva (Spain) | Mar Sorribas Panero e-mail: sorribasm@inta.esPhone: |
Montseny (E) | Institute of Environmental Assessment and Water Research (IDAEA) Consejo Superior de Investigaciones Científicas (CSIC) C/ Jordi Girona 18-26 08034 Barcelona, Spain | Noemi Perez e-mail: noemi.perez@idaea.csic.esPhone: +34 934006100 ext.1521 |
Madrid (E) | Please give me your address. Use the form. | Francisco J Gomez e-mail: fj.gomez@ciemat.es Phone: |
Izana (E)* | Institute: CIAI-AEMET C/La Marina n.20 6ª 38001 Santa Cruz de Tenerife - Spain | Natalia Prats Porta e-mail: npratsp@aemet.es Phone: +34 922151718 |
Checklist | Information | CPC-Status | Information | |
What kind of MPSS do you have? Which integration limits you use? | TSI 3082 | Saturator Temp (°C) | 39.0 | |
Is the Round Robin CPC “status” green? (yes/no) | yes | Condenser Temp (°C) | 23.9 | |
Is the “Data Averaging” set to 60s? (yes/no) | yes | Optics Temp (°C) | 40.0 | |
What data logger is used? (BNC, USB, serial) | USB | Cabinet Temp (°C) | 36.4 | |
Did you take a photo of the setup? (yes/no) | yes | Ambient Press. (kPa) | 92.9 | |
Is there a dryer in the main inlet? (yes/no) | no | Orifice Press. (kPa) | 82.1 | |
What is the flow of the Round Robin CPC. (L/min) | 1.05 | Nozzle Press. (kPa) | 2.5 | |
What is the flow of the MPSS. (L/min) | 1.03 | Laser (mA) | 52 | |
Zero check: Both instruments showed a clear zero? (yes/no) | yes | Liquid level | Full |
Checklist | Information | CPC-Status | Information | |
What kind of MPSS do you have? Which integration limits you use? | TSI 3080 | Saturator Temp (°C) | 39.0 | |
Is the Round Robin CPC “status” green? (yes/no) | yes | Condenser Temp (°C) | 23.9 | |
Is the “Data Averaging” set to 60s? (yes/no) | yes | Optics Temp (°C) | 40.0 | |
What data logger is used? (BNC, USB, serial) | serial | Cabinet Temp (°C) | 33.3 | |
Did you take a photo of the setup? (yes/no) | yes | Ambient Press. (kPa) | 100.4 | |
Is there a dryer in the main inlet? (yes/no) | yes | Orifice Press. (kPa) | 60.7 | |
What is the flow of the Round Robin CPC. (L/min) | 1.041 | Nozzle Press. (kPa) | 3.0 | |
What is the flow of the MPSS. (L/min) | 0.301 | Laser (mA) | 52 | |
Zero check: Both instruments showed a clear zero? (yes/no) | yes | Liquid level | full |
Checklist | Information | CPC-Status | Information | |
What kind of MPSS do you have? Which integration limits you use? | TROPOS N10-800 nm | Saturator Temp (°C) | 39.0 | |
Is the Round Robin CPC “status” green? (yes/no) | yes | Condenser Temp (°C) | 23.9 | |
Is the “Data Averaging” set to 60s? (yes/no) | yes | Optics Temp (°C) | 40.0 | |
What data logger is used? (BNC, USB, serial) | BNC | Cabinet Temp (°C) | 36.9 | |
Did you take a photo of the setup? (yes/no) | no | Ambient Press. (kPa) | 92.5 | |
Is there a dryer in the main inlet? (yes/no) | no | Orifice Press. (kPa) | 77.9 | |
What is the flow of the Round Robin CPC. (L/min) | 1.03 | Nozzle Press. (kPa) | 2.4 | |
What is the flow of the MPSS. (L/min) | 0.98 | Laser (mA) | 53 | |
Zero check: Both instruments showed a clear zero? (yes/no) | yes | Liquid level | full |
Checklist | Information | CPC-Status | Information | |
What kind of MPSS do you have? Which integration limits you use? | Saturator Temp (°C) | |||
Is the Round Robin CPC “status” green? (yes/no) | Condenser Temp (°C) | |||
Is the “Data Averaging” set to 60s? (yes/no) | Optics Temp (°C) | |||
What data logger is used? (BNC, USB, serial) | Cabinet Temp (°C) | |||
Did you take a photo of the setup? (yes/no) | Ambient Press. (kPa) | |||
Is there a dryer in the main inlet? (yes/no) | Orifice Press. (kPa) | |||
What is the flow of the Round Robin CPC. (L/min) | Nozzle Press. (kPa) | |||
What is the flow of the MPSS. (L/min) | Laser (mA) | |||
Zero check: Both instruments showed a clear zero? (yes/no) | Liquid level |
Checklist | Information | CPC-Status | Information | |
What kind of MPSS do you have? Which integration limits you use? | TSI, class3080+CPC3772 (N10-400nm) | Saturator Temp (°C) | 39 | |
Is the Round Robin CPC “status” green? (yes/no) | yes | Condenser Temp (°C) | 23 | |
Is the “Data Averaging” set to 60s? (yes/no) | yes | Optics Temp (°C) | 40 | |
What data logger is used? (BNC, USB, serial) | serial | Cabinet Temp (°C) | 37.1 | |
Did you take a photo of the setup? (yes/no) | yes | Ambient Press. (kPa) | 77.3 | |
Is there a dryer in the main inlet? (yes/no) | no necessary | Orifice Press. (kPa) | 56.2 | |
What is the flow of the Round Robin CPC. (L/min) | 0.975 | Nozzle Press. (kPa) | 2.6 | |
What is the flow of the MPSS. (L/min) | 1.008 | Laser (mA) | 52 | |
Zero check: Both instruments showed a clear zero? (yes/no) | yes | Liquid level | full |
Form for BOX IV - TROPOS