Station
End intercomparison
Information
April 16, 2018
Kosetice (CZ)
May 03, 2018
inter-comparison -> finished
March 29, 2018
Prague (CZ)
April 16, 2018
inter-comparison -> finished
November 14, 2018; 01:10 PM
Vielsalm (B)
November 28, 2018; 01:10 PM
  • Problems with the MPSS on the station -> May -> CPC goes directly to Cabauw
  • May 08 -> call with station to discuss problem -> more checks necessary
  • Oct 24 -> TCPC is traveling to station
  • inter-comparison -> finished November 2018

Cabauw (NL)


Zeppelin (NO)*



Birkenes (NO)*

Station
Address
Contact Person
Kosetice (CZ)
Vladimir Zdimal, Ph.D.
 
head of Department of Aerosols Chemistry and Physics
 
Institute of Chemical Process Fundamentals of the AS CR, v.v.i.
 
Rozvojová 135, CZ-165 02 Praha 6, Czech Republic
Vladimir Zdimal
e-mail: Zdimal@icpf.cas.cz
Phone: +420773400966
Prague (CZ)
Vladimir Zdimal, Ph.D.
 
head of Department of Aerosols Chemistry and Physics
 
Institute of Chemical Process Fundamentals of the AS CR, v.v.i.
 Rozvojová 135, CZ-165 02 Praha 6, Czech Republic
Vladimir Zdimal
 e-mail: Zdimal@icpf.cas.cz
Phone: +420773400966
Vielsalm (B)
Air quality
Institut Scientifique de Service Public (ISSeP)
200, rue du Chera
4000 Liege, Belgium
Benjamin Bergmans
 e-mail: b.bergmans@issep.be
Phone : +32(0)42298218
Cabauw (NL)
Princetonlaan 6, 3584 CB, Utrecht
The Netherlands
Bas Henzing
 e-mail: bas.henzing@tno.nl
Phone:
Alternative contact for technical issues and shipping: marcel.moerman@tno.nl
Zeppelin (NO)*
Please give me your address. Use the form.
Chris Lunder
 e-mail: chris.lunder@nilu.no
Phone:
Birkenes (NO)*
Please give me your address. Use the form.
Chris Lunder
e-mail: chris.lunder@nilu.no
Phone:
Station
Intercomparison time
plots
Information
maintenance
March 22, 2018
CPC-Workshop in TROPOS
Kosetice (CZ)
April 16, 2018 – May 03, 2018


Prague (CZ)
March 29, 2018 – April 16, 2018

maintenance
October 09, 2018
CPC-Workshop in TROPOS



Checklist
Information
CPC-Status
Information
What kind of MPSS do you have?
Which integration limits you use?
TROPOS SMPS
Saturator Temp (°C)
39.0
Is the Round Robin CPC “status” green? (yes/no)
BNC
Condenser Temp (°C)
22.0
Is the “Data Averaging” set to 60s? (yes/no)
yes
Optics Temp (°C)
40.0
What data logger is used?
(BNC, USB, serial)
yes
Cabinet Temp (°C)
27.0
Did you take a photo of the setup? (yes/no)
no
Ambient Press. (kPa)
95.0
Is there a dryer in the main inlet? (yes/no)
no
Orifice Press. (kPa)
74.0
What is the flow of the Round Robin CPC. (L/min)
0.994
Nozzle Press. (kPa)
2.0
What is the flow of the MPSS. (L/min)
1.028
Laser (mA)
-
Zero check: Both instruments showed a clear zero? (yes/no)
yes
Liquid level
ok
Checklist
Information
CPC-Status
Information
What kind of MPSS do you have?
Which integration limits you use?
TSI SMPS 3034 rebuilt at TROPOS
Saturator Temp (°C)
39.0
Is the Round Robin CPC “status” green? (yes/no)
yes
Condenser Temp (°C)
22.0
Is the “Data Averaging” set to 60s? (yes/no)
yes
Optics Temp (°C)
40.0
What data logger is used?
(BNC, USB, serial)
BNC
Cabinet Temp (°C)
28.0
Did you take a photo of the setup? (yes/no)
no
Ambient Press. (kPa)
98.9
Is there a dryer in the main inlet? (yes/no)
yes
Orifice Press. (kPa)
78.4
What is the flow of the Round Robin CPC. (L/min)
0.998
Nozzle Press. (kPa)
2.0
What is the flow of the MPSS. (L/min)
1.015
Laser (mA)
-
Zero check: Both instruments showed a clear zero? (yes/no)
yes
Liquid level
ok
Checklist
Information
TCPC-/ MPSS Status
Information TCPC / MPSS
         start                        end
What kind of MPSS do you have?
Which integration limits you use?
TROPOS MPSS (GUAN WEST 2)
N 10-800 nm
Saturator Temp (°C)
39.0 / 39.0
39.0 / 39.0
Is the Round Robin CPC “status” green? (yes/no)
yes
Condenser Temp (°C)
23.5 / 23.5
22.0 / 22.0
Is the “Data Averaging” set to 60s? (yes/no)
yes
Optics Temp (°C)
40.0 / 40.0
40.0 / 40.0
What data logger is used?
(BNC, USB, serial)
waiting for information
Cabinet Temp (°C)
32.6 / 26.4
31.4 / 26.6
Did you take a photo of the setup? (yes/no)
waiting for information
Ambient Press. (kPa)
93.8 / 95.8
94.8 / 97.1
Is there a dryer in the main inlet? (yes/no)
waiting for information
Orifice Press. (kPa)
62.4 / 64.5
70.0 / 64.6
What is the flow of the Round Robin CPC. (L/min)
1.079-1.094 (14. Nov); 1.058-1.067 (28. Nov)
Nozzle Press. (kPa)
2.7 / 2.7
2.4 / 2.4
What is the flow of the MPSS. (L/min)
1.044-11.050 (14. Nov); 1.029-1.044 (28. Nov)
Laser (mA)
42 / 41
41 / 41
Zero check: Both instruments showed a clear zero? (yes/no)
waiting for information
Liquid level
FULL
FULL
Checklist
Information
CPC-Status
Information
What kind of MPSS do you have?
Which integration limits you use?

Saturator Temp (°C)

Is the Round Robin CPC “status” green? (yes/no)

Condenser Temp (°C)

Is the “Data Averaging” set to 60s? (yes/no)

Optics Temp (°C)

What data logger is used?
(BNC, USB, serial)

Cabinet Temp (°C)

Did you take a photo of the setup? (yes/no)

Ambient Press. (kPa)

Is there a dryer in the main inlet? (yes/no)

Orifice Press. (kPa)

What is the flow of the Round Robin CPC. (L/min)

Nozzle Press. (kPa)

What is the flow of the MPSS. (L/min)

Laser (mA)

Zero check: Both instruments showed a clear zero? (yes/no)

Liquid level

Checklist
Information
CPC-Status
Information
What kind of MPSS do you have?
Which integration limits you use?
Saturator Temp (°C)

Is the Round Robin CPC “status” green? (yes/no)

Condenser Temp (°C)

Is the “Data Averaging” set to 60s? (yes/no)

Optics Temp (°C)

What data logger is used?
(BNC, USB, serial)

Cabinet Temp (°C)

Did you take a photo of the setup? (yes/no)

Ambient Press. (kPa)

Is there a dryer in the main inlet? (yes/no)

Orifice Press. (kPa)

What is the flow of the Round Robin CPC. (L/min)

Nozzle Press. (kPa)

What is the flow of the MPSS. (L/min)

Laser (mA)

Zero check: Both instruments showed a clear zero? (yes/no)

Liquid level

Checklist
Information
CPC-Status
Information
What kind of MPSS do you have?
Which integration limits you use?

Saturator Temp (°C)

Is the Round Robin CPC “status” green? (yes/no)

Condenser Temp (°C)

Is the “Data Averaging” set to 60s? (yes/no)

Optics Temp (°C)

What data logger is used?
(BNC, USB, serial)

Cabinet Temp (°C)

Did you take a photo of the setup? (yes/no)

Ambient Press. (kPa)

Is there a dryer in the main inlet? (yes/no)

Orifice Press. (kPa)

What is the flow of the Round Robin CPC. (L/min)

Nozzle Press. (kPa)

What is the flow of the MPSS. (L/min)

Laser (mA)

Zero check: Both instruments showed a clear zero? (yes/no)

Liquid level


Form for BOX 1 - NILU

Kosetice (CZ)Prague (CZ)Vielsalm (B)Cabauw (NL)Zeppelin (NO)Birkenes (NO)

Contact
If there are changes in the contact table, please write it bellow.




Step 1 - intercomparison period:


------------------------------------------------------------------------
Box status: GGREENG - operational   RREDR - not operational    BBLUEB - maintenance in TROPOS
Box I
Box II
Box III
Box IV
Box V
check in TROPOS
check in TROPOS
check in TROPOS
check in TROPOS
check in TROPOS

©KayWeinhold

last update: July 16, 2019
Back to content